(Figure 38) Field testing of a pressure vacuum breaker involves testing both the internal spring loaded soft seated check valve as well as testing the spring loaded air inlet valve. The testing must be performed with the device pressurized and the air inlet closed. The number 2 shut-off valve must also be closed and the air inlet valve canopy removed.
Method 1 Using a differential pressure gauge
Test 1 Test the internal check valve for tightness of 1 psid in the direction of flow.
With the valve body under pressure, (number 2 shutoff valve closed and number 1 shut-off valve open) bleed test cocks number 1 and number 2.
Hook up the high pressure hose to number 1 test cock and the low pressure hose to number 2 test cock.
Bleed the high pressure hose, and low pressure hose, in that order, and close the test kit needle valves slowly.
Record the differential pressure on the gauge. A reading of 1 psid is acceptable to insure a tight check valve.
Test 2 Test the air inlet valve for a breakaway of 1 psi.
Connect the high pressure hose to test cock number 2, and bleed the high pressure hose.
Shut off number 1 shut-off valve.
Slowly open the bleed valve of the test kit, and observe and record the psi when the air inlet poppet opens. This should be a minimum of 1 psi. Restore the valve to normal service.
Method 2 Using a water column sight tube and 90 degree elbow fitting with bleed needle
Test 1 Test the internal check valve for tightness of 1 psid in the direction of flow.
Assemble sight tube to test cock number 1. Open test cock and fill the tube to a minimum of 36-inches of water height.
Close number 1 shut-off valve.
Open test cock number 2. The air inlet valve should open and discharge water through number 2 test cock.
Open number 1 test cock. The sight tube level of water should drop slowly until it stabilizes. This point should be a minimum of 28-inches of water column which equals 1 psi.
Test 2 Test the air inlet valve for a breakaway of 1 psi.
Assemble sight tube to test cock number 2. Open test cock number 2 and fill the tube to a minimum of 36-inches of water height.
Close number 1 shut-off valve.
Bleed water slowly from the number 2 test cock bleed needle and observe the water column height as it drops.
At the point when the air inlet valve pops open, record the height of the water column. This point should be a minimum of 28-inches of water column which equals 1psi.